More isn't always better: Too much exoskeleton torque can disrupt balance

Abstract

Wearable exoskeletons are a promising tool for augmenting balance and reducing fall risk. Recent work suggests that active ankle exoskeletons need to act faster than the human to improve reactive balance control. However, the magnitude of exoskeleton torque that is best for improving reactive balance remains unknown. Drawing from the optimal torque for minimizing metabolic expenditure, we hypothesized that reactive balance would improve with increased exoskeleton torque. Participants wearing bilateral ankle exoskeletons were instructed to maintain standing balance during 15cm backward support-surface perturbations. Three exoskeleton plantarflexion torque conditions were tested: NO (Off), LOW (15Nm), or HIGH (30Nm). LOW torque improved balance performance compared to NO torque (p<0.001), with a 7{+/-}3% decrease in peak center of mass (CoM) displacement. Although HIGH torque caused a 9{+/-}11% decrease in peak CoM displacement compared to NO torque (p=0.12), it was not significant due to high intersubject variability. Whereas LOW torque decreased peak CoM displacement in all (range: -0.2 to -1.6cm), HIGH torque only decreased it in some (range = 1.2 to -2.6cm). The change in CoM displacement from LOW to HIGH torque was associated with balance ability, quantified by the narrowing beam test (R2=0.29, p=0.06), while this relationship didnt meet conventional statistical significance, likely due to the small sample size, it suggests that higher levels of exoskeleton torque may hinder balance performance in individuals with better balance ability. Taken together, more exoskeleton torque is not always better for balance, highlighting a potential need to personalize exoskeleton torque for balance augmentation.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Han Kim, J., Rastogi, R., Martino, G., Beck, O. N., Shepherd, M. K., Sawicki, G. S., Ting, L. H., Jakubowski, K. L.. 2026-06-03. More isn't always better: Too much exoskeleton torque can disrupt balance. https://doi.org/10.64898/2026.06.02.729541

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related preprints

Wall stiffening is a primary contributor to motility loss in Crohn's disease: an electromechanical modeling study

Fibrotic strictures are among the most disabling complications of Crohn's disease, permanently narrowing the bowel and impairing motility, yet no approved therapy reverses them. Chronic inflammation alters pacemaker-network coupling, smooth-muscle excitability, and calcium-dependent contractility, while fibrosis thickens the bowel wall, narrows the lumen, and changes tissue mechanics. The relative contributions of these coupled electrical, contractile, and structural alterations to motility loss remain unclear. To address this gap, we develop an integrated electromechanical finite-element framework for fibrostenosing Crohn's disease that couples a fibrosis-driven growth model with a FitzHugh-Nagumo electromechanical model. A full-factorial 25 design of experiments is used to quantify the relative effects of electrical diffusivity, excitation threshold, peak active stress, wall stiffness, and hypertrophic remodeling on cyclic lumen-volume deformation. Motility is quantified by the standard deviation of lumen volume over one contraction cycle. Within the parameter ranges examined, increased wall stiffness emerged as the dominant contributor to motility loss, followed by impaired smooth-muscle contractility. Changes in excitation threshold, hypertrophic remodeling, and electrical diffusivity produced substantially smaller effects. Pairwise interactions were small relative to the dominant main effects, indicating that the mechanisms contributed largely through their individual effects. Our findings suggest that limiting wall stiffening while preserving smooth-muscle contractile function may provide a therapeutic strategy for maintaining intestinal motility in fibrostenosing Crohn's disease.

bioengineering↗

Lactate Receptor Activation Alleviates Senescence and Preserves Homeostasis of Aged Arteries

Arteries are among the first tissues to exhibit age-related dysfunction, yet the metabolic mechanisms driving vascular senescence remain poorly understood. Here, analysis of human aortic transcriptomic data identified HCAR1, encoding the lactate receptor GPR81, as one of the genes most significantly downregulated with age. We therefore investigated whether age-associated loss of GPR81 contributes to cellular senescence within the vessel wall. Senescent human endothelial cells and vascular smooth muscle cells accumulated neutral and oxidized lipids and exhibited increased labile iron and ferroptosis. Silencing GPR81 in early-passage cells recapitulated this metabolic phenotype together with multiple hallmarks of cellular senescence. Moreover, endothelial-specific deletion of GPR81 in young mice was sufficient to induce senescent cell accumulation, impaired lipid homeostasis, endothelial dysfunction, and elastin disorganization. Conversely, pharmacological activation of GPR81 with the agonist CHBA restored fatty acid metabolism, promoted glycolytic reprogramming, and attenuated ferroptotic stress and senescence-associated phenotypes. In lamin A knock-in (LAKI) progeroid mice, CHBA reduced arterial lipid accumulation and cellular senescence, shifted vascular cell composition toward a youthful state, improved endothelial integrity, and restored extracellular matrix homeostasis. Together, these findings identify age-associated loss of GPR81 as a driver of vascular metabolic dysfunction and cellular senescence and establish pharmacological GPR81 activation as a promising therapeutic strategy for preserving vascular homeostasis and mitigating age-associated cardiovascular disease.

bioengineering↗

Targeted and bilateral blood flow monitoring in middle cerebral artery using diffuse correlation spectroscopy

Objective: To develop and validate a dual-probe Diffuse Correlation Spectroscopy (DCS) system for non-invasive and simultaneous, monitoring of cerebral blood flow (CBF) in the bilateral Middle Cerebral Artery (MCA) territories, and expanding the utility of conventional DCS limited to cortical-volume-based CBF measurements to vessel-specific cerebral perfusion monitoring. Methods: A dual-probe DCS system was designed for non-invasive monitoring of MCA-specific perfusion. Probe placement and protocol optimization study has been performed using anatomical landmarks, motor and speech activation tasks in healthy volunteers. System stability and repeatability were further evaluated in a pilot cohort of 30 healthy (age, 25{+/-}7 years) participants using optimized probe position and protocol. A bilateral MCA ischemic Lacunar Infract stroke case report also validated the feasibility of the system in clinical settings. Results: Measurements demonstrated superior sensitivity towards MCA-territory perfusion at targeted probe locations compared to off-MCA positions. In pilot cohort, significant increase of 30.34 {+/-} 21.56% and 36.48 {+/-} 21.22% in rCBF corresponding to hand squeeze and speech task respectively showed reproducible physiological responsiveness of the system (p<0.001). Measurement done on a patient with bilateral MCA ischemic Lacunar Infract stroke showed a significant change of 30% during speech for both the MCAs but no significant change is observed for hand squeeze tasks. Conclusion: The custom built dual-probe DCS system enables non-invasive, operator-independent, targeted and continuous monitoring of rCBF within bilateral MCA territories. Significance: This approach enables the potential use of DCS system for bilateral and vessel-specific monitoring of cerebral perfusion in the MCA territories.

bioengineering↗