Search bioRxiv⌕ Search

bioRxiv · 10.64898/2026.10.26.761989

Three-Dimensional Fluid-Structure Interaction Reveals Trade-offs Among Oxygen Delivery, Shear, and Washout in Pediatric Roller-Pump Cardiopulmonary Bypass

Abstract

Objective: To determine how roller speed, tubing occlusion, and hematocrit jointly regulate oxygen delivery, shear, and washout in pediatric roller-pump cardiopulmonary bypass. Methods: A three-dimensional, closed-loop fluid-structure interaction model resolved roller-tube contact, tubing deformation, and blood flow. Forty simulations combined 4 roller speeds (38-95 revolutions per minute), 2 occlusion levels (77% and 83%), and 5 hematocrit levels (15%-50%). Cycle-averaged flow, oxygen delivery, peak shear stress, and viscosity-normalized relative residence time were quantified. Results: Roller compression produced retrograde leakage through the occluded gap, localized peak shear stress exceeding 100 pascals, and adjacent stagnation and recirculation. Flow increased nearly linearly with roller speed and was higher at greater occlusion, with little hematocrit dependence. Higher speed and occlusion increased peak shear stress while reducing relative residence time; higher hematocrit further increased peak shear stress. Similar oxygen delivery could arise from substantially different flow environments: among 10 conditions providing oxygen delivery of approximately 9-12 milliliters per minute, peak shear stress ranged from approximately 25 to 70 pascals. Within conditions satisfying an illustrative minimum oxygen-delivery threshold, lower hematocrit permitted speed and occlusion settings yielding lower peak shear stress and relative residence time. Conclusions: Pediatric roller-pump settings that provide similar oxygen delivery can expose blood to markedly different shear and washout environments. Joint evaluation of these metrics may guide selection of speed, occlusion, and hematocrit to meet a target oxygen delivery while reducing shear and improving washout. Keywords: Cardiopulmonary bypass; roller pump; pediatric cardiac surgery; fluid-structure interaction; shear stress; relative residence time

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Tu, Y., Yeh, Y.-T., de Lecea, A., Charette, K., Bohuta, L., Nigam, V., del Alamo, J. C., Greene, C. L.. 2026-10-28. Three-Dimensional Fluid-Structure Interaction Reveals Trade-offs Among Oxygen Delivery, Shear, and Washout in Pediatric Roller-Pump Cardiopulmonary Bypass. https://doi.org/10.64898/2026.10.26.761989

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related preprints

Mesoscale Soft Robotics for Mechanically Perturbing and Functionally Monitoring Responses in Human Organoids

Human organoids represent promising models for studies of significance to the biology and medical communities. Of the existing approaches for monitoring organoids, none offers the ability to apply programmable mechanical forces while simultaneously recording multi-modal physiological responses, of relevance to phenomena spanning from traumatic brain injuries to cardiac rhythm disorders. Here, we present a mesoscale soft robotic organoid interface that enables programmable delivery of forces to organoids of diverse shapes, and to map electrophysiological activity and mechanical responses. The platform combines actuation elements with integrated sensors for quantitative measurement and control. Studies using this technology reveal that spheroidal cortical organoids exhibit distinct responses to mechanical trauma, including silencing, recovery, and changes in neuronal firing patterns. Additional experiments on cardiac ring tissues demonstrate that sustained stretching induces beat-to-beat rhythm variability and altered conduction. The results establish a framework for investigating force-triggered transitions between physiological and disease-like states in human tissue models.

bioengineering↗

Few-shot regulatory sequence design with a genomic language model

A central challenge in generative biology is translating knowledge learned from natural sequences and experiments into designs with desired functions. In-context learning (ICL) is an emergent capability of large language models that allows them to adapt to new tasks from a few examples without additional training, enabling these models to become general-purpose learners. Here, we show that the genome language model, Evo 2, exhibits generative ICL in biological sequence design. Prompting with just tens of examples conditions Evo 2 to generate novel and diverse regulatory DNA sequences without any additional training. Mechanistic analysis suggests that genomic ICL may originate from induction, similar to phenomena observed in natural language. Reinforcement learning leveraging sequence-to-function models as experiment-derived rewards further strengthens ICL. We first validate this approach by generating enhancers that drive cell-type-specific gene expression in three human cell lines. Extending this approach in vivo, enhancers designed using human regulatory element data selectively label layer 5 extratelencephalic neurons in the mouse cortex with high specificity and stronger activity than existing enhancers. Our study provides evidence of emergent properties in genomic sequence models and introduces a promising approach for improving functional conditioning in generative biology.

bioengineering↗

IOP-independent treatment for neurodegeneration in glaucoma using optimized ROS-scavenging microparticles for sustained release of erythropoietin

Glaucoma is a slow-progressing neurodegenerative disease of the optic projection to the brain and a leading cause of blindness worldwide. Current treatments are targeted towards lowering intraocular pressure (IOP), the only known, modifiable risk factor; however, this strategy is ineffective in many patients. An early hallmark of glaucomatous neurodegeneration is increased reactive oxygen species (ROS), representing an alternative opportunity for intervention to halt or slow disease progression. In this work, we optimize a ROS-scavenging copolymer, poly(propylene sulfide-co-ethylene sulfide) (PPSES), for formulation of size-controlled, stabilized, antioxidant microparticles (MPs) for sustained release of the protein erythropoietin (EPO). EPO promotes biological antioxidant defense via activation of NRF2-mediated transcription of antioxidant genes. This MP system overcomes the short half-life of EPO and leverages the combined benefits of bioactive EPO with long-lasting, inherently antioxidant PPSES MPs. The lead EPO-MP formulation preserves neurons in both mouse and non-human primate models of glaucoma through suppression of oxidative stress and optic nerve degeneration. EPO-loaded PPSES MPs represent a translationally relevant therapy for long-term management of glaucoma independent of IOP.

bioengineering↗