APRIL: Adaptive Regression-Based Two-Dimensional Quantitative Anisotropy Imaging Using Acoustic Radiation Force Impulse

Abstract

ObjectiveThis study aims to develop and validate a quantitative, depth-resolved anisotropy imaging framework that extends ARFI-based focal degree-of-anisotropy (DoA) estimation into two-dimensional mapping by modeling the depth-dependent relationship between shear modulus ratio (SMR) and peak displacement ratio (PDR). MethodsWe propose APRIL (Adaptive Polynomial Regression for anisotropy Imaging via ARFI-induced DispLacements), a framework for quantitative, depth-resolved DoA imaging that adaptively selects polynomial regression or shape-preserving spline interpolation based on excitation PSF asymmetry. Training data were generated using an LS-DYNA3D + Field II simulation pipeline in homogeneous transversely isotropic media (SMR 0.9-4.9). Testing included shifted SMRs under varied acoustic conditions and three heterogeneous inclusion configurations (anisotropic inclusion in isotropic background and vice versa). Experimental validation was performed in an in-vivo murine tumor model over the time, ex-vivo chicken breast, and tissue-mimicking gelatin phantoms, using a Verasonics system with an L11-5v transducer. ResultsAPRIL achieved depth-resolved SMR prediction errors below 9% over 10-30 mm, with highest accuracy in the focal region (MAE 2.3%, RMSE < 0.1) and stable performance across PSF transition zones. In heterogeneous phantoms, it reconstructed anisotropy maps with SSIM up to 86% and MPE below 7%, accurately delineating inclusion boundaries. Under acoustic parameter variations, mean absolute errors remained below 10%, demonstrating robustness to system and tissue heterogeneity. ConclusionAPRIL enables robust, two-dimensional anisotropy imaging beyond focal estimates. SignificanceThe method provides a physically grounded and generalizable framework for clinically viable anisotropy biomarkers in muscle, tendon, kidney, tumor and breast tissues. HighlightsO_LINovelty: APRIL introduces LoA-conditioned adaptive polynomial-spline regression to extend ARFI-based anisotropy estimation from focal point estimates into full 2D depth-resolved SMR imaging. C_LIO_LIResults: APRIL achieved SMR prediction errors below 9% over 10-30 mm, SSIM up to 86% in heterogeneous phantoms, MAE below 10% under acoustic variations, tracked tumor anisotropy progression in vivo, and differentiated anisotropic inclusion versus isotropic background in tissue-mimicking gelatin phantom. C_LIO_LISignificance: APRIL enables clinically viable, spatially resolved anisotropy biomarker imaging in muscle, tendon, kidney, and tumor tissues without requiring heterogeneous training data. C_LI

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Hassan, M. W., Crook, K., Gi, Y. J., Lee, J., Hossain, M. M.. 2026-07-01. APRIL: Adaptive Regression-Based Two-Dimensional Quantitative Anisotropy Imaging Using Acoustic Radiation Force Impulse. https://doi.org/10.64898/2026.06.30.735710

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related preprints

Dynamic Compression Platform for Live Imaging of Scaffold-Transmitted Cellular Mechanoresponses

Mechanical characterization of biomaterial scaffolds is essential to evaluate their capacity to meet the functional demands of target tissues in tissue engineering and regenerative medicine applications. Scaffolds designed to interface with living tissues must support the transmission of mechanical cues to resident cells and stimulate mechanosignaling pathways that are essential to their function. In joints, bone and cartilage cells act as primary mechanosensors, converting mechanical stimuli into biochemical signals that regulate tissue homeostasis and remodelling. Therefore, evaluating cellular mechanoresponses to scaffold-transmitted compression in vitro can inform the development of functional tissue-engineered constructs. For example, poly({epsilon}-caprolactone) (PCL) scaffolds are highly relevant for bone and cartilage tissue engineering due to their biocompatibility, stable mechanical properties and slow degradation. Here, we applied a custom-built device to study compression-induced mechanosignaling in MC3T3-E1 pre-osteoblast cells. The device is composed of a polydimethylsiloxane (PDMS) pillar, a force-sensing load cell, and a piezoelectric linear track. A protocol is described in which MC3T3-E1 cells are repeatedly compressed, while in parallel live tracking of force measurements and live imaging of intracellular calcium dynamics in MC3T3-E1 cells are recorded. PCL scaffolds fabricated by melt electrowriting (MEW) were subsequently integrated into the platform. Scaffold-transmitted compression triggered dynamic increases in cytosolic calcium; in MC3T3-E1 cells located directly under the PCL microfibers, but also in cells located in the interfiber spaces. This device and workflow facilitate in vitro investigations of real-time cellular mechanoresponses to dynamic compression applied with biomaterial scaffolds, and provides a testing platform for evaluating the mechanotransductive properties of scaffolds intended for tissue engineering applications.

bioengineering↗

Ultrasound Tracking Reveals Progressive Regional Strain Differences in Human Achilles Tendons During Fatigue Loading

Ultrasound is commonly used to assess structural changes in symptomatic Achilles tendons, but quantitative biomechanical metrics for progressive tendon deterioration remain limited. The goal of this study was to develop and validate an automated ultrasound tracking algorithm for regional tendon deformation and evaluate strain progression in survived and ruptured tendons during fatigue loading. We hypothesized that maximum strain, average strain, and strain heterogeneity would exhibit different trajectories between groups. Ten cadaveric Achilles tendons underwent cyclic loading with stress tests every 500 cycles until rupture or 150,000 cycles. Ultrasound images acquired during stress tests were analyzed using an automated tracking algorithm to generate spatially resolved regional strain fields. Ultrasound-derived bulk strain was highly correlated with actuator-derived strain in survived (R^2 = 0.968 +/- 0.017) and ruptured tendons (R^2 = 0.972 +/- 0.014). Maximum and average longitudinal strains progressively diverged between groups across fatigue life (Group x FatigueLife: p = 0.003 and p < 0.0001, respectively). During the first 10,000 cycles, average strain decreased in survived tendons ({beta} = -0.0268%, p = 0.0215) but not ruptured tendons ({beta} = 0.0147%, p = 0.1197), with a significant Group x Cycle interaction (p = 0.0061). This study demonstrates that the algorithm quantified Achilles tendon deformation with high fidelity and enabled spatially resolved strain assessment throughout fatigue loading. Maximum and average strain followed different trajectories between groups, whereas strain heterogeneity did not. Early differences in tendon biomechanics suggest that regional strain behavior may change before pronounced differences in absolute magnitude develop.

bioengineering↗

Brain organoid computing for robotic decision-making

Biomimicry has inspired the evolution of robotics toward greater autonomy, adaptability, and symbiosis with humans and dynamic environments. However, current robotic systems still face major challenges in recapitulating the high-efficiency decision-making capabilities of the human brain under complex and dynamic conditions. Here, we present Brainobot, a biohybrid robotic system that establishes a brain organoid controller as a high-level robotic decision-making layer for closed-loop embodiment. By leveraging brain organoid reservoir computing, Brainobot interacts with dynamic environments by receiving and processing sensory inputs and generating motor actions. As a proof-of-concept demonstration, Brainobot is implemented in a humanoid robotic system to perform real-world tasks, including object grasping and laser chasing. Interestingly, Brainobot exhibits unique features, including cross-task adaptivity, high computing efficiency, and low energy consumption. Thus, our approach may provide insights for advancing robotic embodiment and understanding biological decision-making.

bioengineering↗