Characterization of real B0 shim fields generated by higher order B0 shim systems of whole body human 3T and 7T MRI systems

Abstract

PurposeEnsuring optimal homogeneity of the static magnetic field (B0) is critically important for significantly enhancing the image quality, spectral resolution, and overall diagnostic accuracy of magnetic resonance imaging (MRI) and magnetic resonance spectroscopy/spectroscopic imaging (MRS/MRSI), especially at high and ultra-high field strengths. Achieving this goal relies on the deployment of advanced, higher-order shim hardware, which is indispensable for effective and precise B0 shimming. MethodsIn this study, we acquired B0 field maps to characterize the performance of each shim channel of vendor-provided B0 shim hardware integrated into whole body human 7T MRI system. To rigorously evaluate and compare the B0 field homogeneity achievable using these devices, we utilized spherical harmonic expansions to fit the measured B0 distributions for each shim channel in each of the investigated built-in B0 shim systems. This methodology enabled a systematic assessment of shim field "purity" and the effectiveness of different B0 shim configurations. ResultsWhile our analysis confirmed that the linear-order shim terms provided by the gradient coils maintained high purity, certain second-order shim terms and all third-order B0 shim coils of all vendors 7T whole body human MRI systems exhibit a high level of impurity and more specifically introduce substantial linear field components. These unintended linear contributions interfere with the linear shim fields provided by the gradient coils negating their corrective influence. As a result, the 3rd order B0 shim coils and some 2nd order B0 shim coil elements compromised the overall effectiveness of the B0 shimming process rather than improving it. Although pure higher order B0 shim fields were previously shown to substantially improve B0 field inhomogeneity, the vendor implemented versions of higher order B0 shim systems in current 7T whole-body human MRI scanners cannot achieve the desired level of uniformity and partly perform worse than first order B0 shimming. Substantial deviations between 2nd order and especially 3rd order real shim fields versus ideal shim fields demonstrate the necessity of real shim field calibration and call for improvement of future 2nd and 3rd order B0 shim coil design for human whole-body MRI scanners. Our findings offer valuable guidance for optimizing B0 shimming strategies at high and ultra-high field strengths, ultimately enhancing image quality. ConclusionRecent evaluations have demonstrated that certain second- and third-order shim coils, as implemented in current commercial 7T human whole-body MRI systems, introduce unintended lower-order (linear) field components. These field "impurities" substantially deviate from the desired higher-order spatial harmonics, thereby diminishing the orthogonality of the higher order B0 shim system. This in turn largely compromises the effectiveness of higher-order B0 shimming at whole-body 7T MRI scanners. These findings emphasize the necessity of conducting accurate shim field calibration to best utilize the current 7T B0 shim system configurations on one hand. On the other hand, improving the purity of these commercial B0 shim systems in the design phase to ensure that the higher-order terms accurately match their intended spherical harmonic profiles, particularly for the third-order shim terms--would largely enhance the achievable B0 field homogeneity at 7T. This improvement has directly clinically significant outcomes.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Jani, M., Su, S., Gast, L., Nagel, A., Henning, A.. 2025-08-27. Characterization of real B0 shim fields generated by higher order B0 shim systems of whole body human 3T and 7T MRI systems. https://doi.org/10.1101/2025.08.26.672172

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related preprints

RNASeek: A Cross-Phyla Generative Foundation Model for Multipurpose RNA Modeling and Reinforcement Learning-Based Design

RNA plays central roles in regulating information flow and provides a versatile substrate for engineering biological functions. While large language models (LLMs) have transformed natural language processing and protein design, a general framework connecting RNA foundation models to functional sequence design remains limited. Here, we present RNASeek, a 1.6-billion-parameter generative foundation model built on a DeepSeek architecture and trained on a cross-phyla transcriptomic corpus for RNA sequence representation and generation. Natural-language tokens enable flexible conditional prediction and sequence design using a unified backbone. RNASeek captures species-specific transcript features and intron-exon boundaries in a zero-shot setting. We then fine-tune RNASeek to predict ribozyme self-cleavage activity and viral mRNA stability, revealing interpretable sequence features associated with function, including ribozyme loop flexibility and stem stability, as well as AU-rich motifs associated with mRNA stability. We use these functional predictors as reward models and apply Group Relative Policy Optimization (GRPO) to update the generation policy of RNASeek toward sequences with desired properties. GRPO-guided generation produces faster-cleaving ribozymes and stability-enhancing 3' UTRs while satisfying user-specified IUPAC constraints. Experimentally validated RNASeek-generated ribozymes achieve wild-type levels of activity, while RNASeek-generated 3' UTR sequences exceed the performance of the training data and benchmarked AI-generated 3' UTRs. Together, RNASeek establishes a unified pretrain-predict-optimize framework that connects learned RNA function to controllable de novo sequence design and provides a general strategy for engineering regulatory RNAs with desired properties.

bioengineering↗

Joint Vector Flow Mapping and Segmentation: Ill-Posedness,Differentiable Bayesian Inference, and Synthetic Vortex-FlowBenchmarks

Vector flow mapping (VFM) reconstructs left-ventricular (LV) blood velocity from color-Doppler echocardiography by combining the measured beamwise component with physical and regularizing constraints. Analysis of the discrete VFM formulation shows that the inverse problem is intrinsically ill posed: the occurrence of singular modes can be predicted from the geometry of the segmented blood-pool domain, the imposed boundary conditions, and the degree of smoothing. These modes can propagate uncertainty along entire transverse bands of the reconstructed velocity field, yet conventional VFM neither quantifies this uncertainty nor allows for correcting the blood-pool segmentation. We introduce Bayesian VFM (B--VFM), a hierarchical framework that jointly infers radial and transverse velocities, a probabilistic blood-pool mask, their spatially resolved uncertainties, and hyperparameters weighting Doppler and segmentation fidelity, mass conservation, boundary conditions, and smoothness. The discretized posterior admits a closed-form gradient and exact Hessian, enabling computationally efficient, gradient-based MAP estimation, sampling, and direct analysis of ill-posed modes. Posterior inference combines Gibbs sampling of conjugate Gamma-distributed hyperparameters with conditional maximum-a-posteriori estimation and a Laplace approximation for the high-dimensional velocity and mask fields. To accommodate systematic departures from planar mass conservation, B-VFM can learn the covariance of the planar divergence residual from an ensemble of flows and incorporate it as a structured model-discrepancy prior. Independent chains converged reproducibly, while covariance priors learned from flow ensembles illustrated how model discrepancies can be incorporated into the inference. B--VFM was evaluated using Lamb-Chaplygin dipoles under ideal conditions and with Doppler corruption, Doppler voids, and segmentation defects, and using the Hicks-Moffatt family of spherical vortices to assess violations of planar mass conservation. The method produced smooth reconstructions, localized uncertainty near unreliable measurements and regions of model inconsistency, and used flow information to correct segmentation errors. Within the tested vortex family, the data-informed planar divergence prior reduced velocity bias and mask distortion. B--VFM thus provides an uncertainty-aware reconstruction method and a flexible foundation for future VFM formulations incorporating additional priors, observations, and physical models. Future work will evaluate the method using clinical data and more complex three-dimensional benchmark flows.

bioengineering↗

Computational design of a versatile, zero-radius proximity labeling enzyme

The ability to map protein interactomes and organelle proteomes is foundational for achieving a molecular understanding of living cells. Proximity labeling (PL) provides a powerful strategy for this, but existing enzymes and photocatalysts are limited by their spatial resolution, reliance on biotin, and/or in vivo compatibility. Here we report FlexID, an engineered promiscuous ligase that catalyzes the rapid attachment of diverse small-molecule probes to proximal endogenous proteins. Critically, FlexID operates through a zero-radius, direct-contact mechanism, offering superior spatial precision compared to existing PL tools. We engineered FlexID by combining the strengths of sequence- and structure-trained computational models to enhance its catalytic activity and structural stability. Biophysical analysis revealed that specific conformational changes in FlexID improve its ability to recognize diverse target proteins while simultaneously preventing the premature release of the reactive intermediate. We demonstrate FlexID's versatility through in vivo proximity labeling, comprehensive organelle proteome mapping, and a high-throughput, fluorescence-based screen for molecular glues. Our work shows that computational methods can be harnessed to create mechanistically distinct PL enzymes and establishes FlexID as a flexible, high-resolution tool for mapping protein interactions and proteomes in living cells.

bioengineering↗