Design Guidelines for Dry Electrode Tip Geometry in Electroencephalography Measurements: A Proposal Based on the Scalp's Mechanical Response

Abstract

In electroencephalography (EEG) measurements using dry electrodes, a trade-off between signal stability and user comfort is a critical barrier to long-term, wearable applications. While various approaches have been proposed to address this issue, the mechanical impact of electrode tip geometry has not been adequately quantified, and most existing evaluations predominantly rely on subjective assessments. To address this gap with a quantitative, mechanics-based framework, the current study aimed to identify an optimized electrode tip geometry that minimizes mechanical stress on the scalp even under tilted contact conditions. Finite element analysis was conducted using strain energy density (SED)-- a mechanical index known to correlate with neural impulse activity--as a quantitative indicator of the mechanical influence of tip geometry on the skin. Six types of electrode tip geometries, ranging from flat to hemispherical, were defined based on the ratio of fillet radius to prong radius. These geometries were analyzed under inclination angles from 0{degrees} to 5{degrees}, and their peak SED values were compared. Additionally, a geometry optimization using an iterative search algorithm was performed to minimize peak SED under the 5{degrees} tillt. The findings revealed that intermediate fillet geometries with gently rounded edges more effectively reduce peak SED under inclined conditions. Optimization further identified a geometry ratio of [Formula] as the most effective tip geometry for minimizing mechanical loading under the specified conditions. These results offer a potential geometric design guideline for dry EEG electrodes that can help maintain user comfort across varying inclination angles.

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Araki, S., Nakatani, S., Araki, N.. 2025-08-01. Design Guidelines for Dry Electrode Tip Geometry in Electroencephalography Measurements: A Proposal Based on the Scalp's Mechanical Response. https://doi.org/10.1101/2025.07.30.667786

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related preprints

Dynamic Compression Platform for Live Imaging of Scaffold-Transmitted Cellular Mechanoresponses

Mechanical characterization of biomaterial scaffolds is essential to evaluate their capacity to meet the functional demands of target tissues in tissue engineering and regenerative medicine applications. Scaffolds designed to interface with living tissues must support the transmission of mechanical cues to resident cells and stimulate mechanosignaling pathways that are essential to their function. In joints, bone and cartilage cells act as primary mechanosensors, converting mechanical stimuli into biochemical signals that regulate tissue homeostasis and remodelling. Therefore, evaluating cellular mechanoresponses to scaffold-transmitted compression in vitro can inform the development of functional tissue-engineered constructs. For example, poly({epsilon}-caprolactone) (PCL) scaffolds are highly relevant for bone and cartilage tissue engineering due to their biocompatibility, stable mechanical properties and slow degradation. Here, we applied a custom-built device to study compression-induced mechanosignaling in MC3T3-E1 pre-osteoblast cells. The device is composed of a polydimethylsiloxane (PDMS) pillar, a force-sensing load cell, and a piezoelectric linear track. A protocol is described in which MC3T3-E1 cells are repeatedly compressed, while in parallel live tracking of force measurements and live imaging of intracellular calcium dynamics in MC3T3-E1 cells are recorded. PCL scaffolds fabricated by melt electrowriting (MEW) were subsequently integrated into the platform. Scaffold-transmitted compression triggered dynamic increases in cytosolic calcium; in MC3T3-E1 cells located directly under the PCL microfibers, but also in cells located in the interfiber spaces. This device and workflow facilitate in vitro investigations of real-time cellular mechanoresponses to dynamic compression applied with biomaterial scaffolds, and provides a testing platform for evaluating the mechanotransductive properties of scaffolds intended for tissue engineering applications.

bioengineering↗

Ultrasound Tracking Reveals Progressive Regional Strain Differences in Human Achilles Tendons During Fatigue Loading

Ultrasound is commonly used to assess structural changes in symptomatic Achilles tendons, but quantitative biomechanical metrics for progressive tendon deterioration remain limited. The goal of this study was to develop and validate an automated ultrasound tracking algorithm for regional tendon deformation and evaluate strain progression in survived and ruptured tendons during fatigue loading. We hypothesized that maximum strain, average strain, and strain heterogeneity would exhibit different trajectories between groups. Ten cadaveric Achilles tendons underwent cyclic loading with stress tests every 500 cycles until rupture or 150,000 cycles. Ultrasound images acquired during stress tests were analyzed using an automated tracking algorithm to generate spatially resolved regional strain fields. Ultrasound-derived bulk strain was highly correlated with actuator-derived strain in survived (R^2 = 0.968 +/- 0.017) and ruptured tendons (R^2 = 0.972 +/- 0.014). Maximum and average longitudinal strains progressively diverged between groups across fatigue life (Group x FatigueLife: p = 0.003 and p < 0.0001, respectively). During the first 10,000 cycles, average strain decreased in survived tendons ({beta} = -0.0268%, p = 0.0215) but not ruptured tendons ({beta} = 0.0147%, p = 0.1197), with a significant Group x Cycle interaction (p = 0.0061). This study demonstrates that the algorithm quantified Achilles tendon deformation with high fidelity and enabled spatially resolved strain assessment throughout fatigue loading. Maximum and average strain followed different trajectories between groups, whereas strain heterogeneity did not. Early differences in tendon biomechanics suggest that regional strain behavior may change before pronounced differences in absolute magnitude develop.

bioengineering↗

Brain organoid computing for robotic decision-making

Biomimicry has inspired the evolution of robotics toward greater autonomy, adaptability, and symbiosis with humans and dynamic environments. However, current robotic systems still face major challenges in recapitulating the high-efficiency decision-making capabilities of the human brain under complex and dynamic conditions. Here, we present Brainobot, a biohybrid robotic system that establishes a brain organoid controller as a high-level robotic decision-making layer for closed-loop embodiment. By leveraging brain organoid reservoir computing, Brainobot interacts with dynamic environments by receiving and processing sensory inputs and generating motor actions. As a proof-of-concept demonstration, Brainobot is implemented in a humanoid robotic system to perform real-world tasks, including object grasping and laser chasing. Interestingly, Brainobot exhibits unique features, including cross-task adaptivity, high computing efficiency, and low energy consumption. Thus, our approach may provide insights for advancing robotic embodiment and understanding biological decision-making.

bioengineering↗