bioRxiv · 10.1101/2023.08.11.553060
Correlation of two different devices for the evaluation of primary implant stability depending on dental implant length and bone density: an in vitro study
Abstract
Non-invasive objective implant stability measurements are needed to determine the appropriate timing of prosthetic fitting after implant placement. We compared the early implant stability results obtained using resonance frequency analysis (RFA) and damping capacity analysis (DCA) depending on the implant length and bone density. Total 60, 4.0 mm diameter implants of various lengths (7.3 mm, 10 mm, and 13 mm) were used. Group I and Group II had implants placed in an artificial bone model with a uniform density of 15 PCF (0.24 g/cm3) and 30 PCF (0.48 g/cm3), respectively. RFA was performed using an Osstell(R) Beacon+; DCA was performed using Anycheck(R). Measurements were repeated five times for each implant. Statistical significance was set at P <0.05. In Group I, bone density and primary implant stability were positively correlated, while implant length and primary implant stability were positively correlated. In Group II, the implant stability quotient (ISQ) and implant stability test (IST) values in did not change significantly above a certain length. Primary implant stability was positively correlated with bone density and improved with increasing implant length at low bone densities. Compared with the Osstell(R) Beacon+, the simplicity of Anycheck(R) was easy to use and accessible.
Source connections
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Lee, J., Lim, Y.-J., Kim, B., Ahn, J.-S., Baek, Y.-W., Lim, B.. 2023-08-15. Correlation of two different devices for the evaluation of primary implant stability depending on dental implant length and bone density: an in vitro study. https://doi.org/10.1101/2023.08.11.553060
Cite the original work for its findings. Save a collection to share your selection of sources.