Abstract
Intracortical microstimulation (ICMS) enables applications ranging from neuroprosthetics to causal circuit manipulations. However, the resolution, efficacy, and chronic stability of neuromodulation is often compromised by the adverse tissue responses to the indwelling electrodes. Here we engineer ultraflexible stim-Nanoelectronic Threads (StimNETs) and demonstrate low activation threshold, high resolution, and chronically stable ICMS in awake, behaving mouse models. In vivo two-photon imaging reveals that StimNETs remain seamlessly integrated with the nervous tissue throughout chronic stimulation periods and elicit stable, focal neuronal activation at low currents of 2 A. Importantly, StimNETs evoke longitudinally stable behavioral responses for over eight months at markedly low charge injection of 0.25 nC/phase. Quantified histological analysis show that chronic ICMS by StimNETs induce no neuronal degeneration or glial scarring. These results suggest that tissue-integrated electrodes provide a path for robust, long-lasting, spatially-selective neuromodulation at low currents which lessen risks of tissue damage or exacerbation of off-target side-effects.
Source connections
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Lycke, R., Kim, R., Zolotavin, P., Montes, J., Koszeghy, A., Altun, E., Noble, B., Yin, R., He, F., Totah, N., Xie, C., Luan, L.. 2023-02-21. Low-threshold, high-resolution, chronically stable intracortical microstimulation by ultraflexible electrodes. https://doi.org/10.1101/2023.02.20.529295
Cite the original work for its findings. Save a collection to share your selection of sources.