Abstract

BackgroundSince the advent of magnetic resonance imaging (MRI) nearly four decades ago, there has been a quest for ever-higher magnetic field strengths. Strong incentives exist to do so, as increasing the magnetic field strength increases the signal-to-noise ratio of images. However, ensuring patient safety becomes more challenging at high and ultrahigh field MRI (i.e., [≥]3 T) compared to lower fields. The problem is exacerbated for patients with conductive implants, such as those with deep brain stimulation (DBS) devices, as excessive local heating can occur around implanted lead tips. Despite extensive effort to assess radio frequency (RF) heating of implants during MRI at 1.5 T, a comparative study that systematically examines the effects of field strength and various exposure limits on RF heating is missing. PurposeThis study aims to perform numerical simulations that systematically compare RF power deposition near DBS lead models during MRI at common clinical and ultra-high field strengths, namely 1.5, 3, 7, and 10.5 T. Furthermore, we assess the effects of different exposure constraints on RF power deposition by imposing limits on either the B1+ or global head specific absorption rate (SAR) as these two exposure limits commonly appear in MRI guidelines. MethodsWe created 33 unique DBS lead models based on postoperative computed tomography (CT) images of patients with implanted DBS devices and performed electromagnetic simulations to evaluate the SAR of RF energy in the tissue surrounding lead tips during RF exposure at frequencies ranging from 64 MHz (1.5 T) to 447 MHz (10.5 T). The RF exposure was implemented via realistic MRI RF coil models created based on physical prototypes built in our institutions. We systematically examined the distribution of local SAR at different frequencies with the input coil power adjusted to either limit the B1+ or the global head SAR. ResultsThe MRI RF coils at higher resonant frequencies generated lower SARs around the lead tips when the global head SAR was constrained. The trend was reversed when the constraint was imposed on B1+. ConclusionAt higher static fields, MRI is not necessarily more dangerous than at lower fields for patients with conductive leads. Specifically, when a conservative safety criterion, such as constraints on the global SAR, is imposed, coils at a higher resonant frequency tend to generate a lower local SAR around implanted leads due to the decreased B1+ and, by proxy, E field levels.

Source connections

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Kazemivalipour, E., Sadeghi-Tarakameh, A., Keil, B., Eryaman, Y., Atalar, E., Rad, L. G.. 2022-12-15. Effect of field strength on RF power deposition near conductive leads: A simulation study of SAR in DBS lead models during MRI at 1.5 T - 10.5 T. https://doi.org/10.1101/2022.12.13.520303. https://searchbiorxiv.org/record/biorxiv%3A10.1101%2F2022.12.13.520303/effect-of-field-strength-on-rf-power-deposition-near-conductive-leads-a-simulation-study-of-sar-in-d

Citations retain the original authors and DOI and include this site’s record URL. If this discovery service contributed to your research, you can also cite Search bioRxiv by Indiciae: BibTeX · RIS.

KEEP EXPLORING

Related preprints

Conformal Quadrature RF Coils with a Semi-open Architecture for 9.4 T Brain MRI from Mice to Macaques

Objective: To develop a semi-open quadrature RF coil architecture that combines brain coverage with low interport coupling at 9.4 T across animal sizes. Methods: Head-conformal high-pass resonators with an adjustable peripheral branch were constructed for four animal head sizes. A shared-rung circuit model, two-capacitance electromagnetic screening, and accepted-power-normalized field analysis guided the implementation. Electromagnetic simulations, loaded bench measurements, saline-phantom comparison with a quadrature birdcage, and ex vivo coronal brain SNR analysis evaluated the macaque implementation. Results: The peripheral branch reduced loaded interport transmission from approximately -10 to & -35 dB. Saline-phantom raw SNR medians were 298.9 and 257.8 for the conformal and birdcage coils, respectively (+16.0%). Voxel-normalized pooled ex vivo brain SNR was 648.7 versus 552.8 (+17.3%). Structural imaging demonstrated imaging feasibility in mice, rats/tree shrews, a canine specimen, and macaques, including in vivo macaque imaging with voxel dimensions of 176 x 156 x 156 micrometres along the three spatial dimensions. Conclusion: The architecture accommodated different animal head sizes through size-specific adjustment while maintaining low interport coupling. Significance: The controlled macaque-scale comparisons support a low-channel-count transceiver option for the demonstrated structural MRI tasks.

biophysics↗

A single LIM domain is sufficient for strain-selective binding of zyxin to F-actin

Cells sense mechanical cues through force-dependent changes in protein binding and activity, and the actin cytoskeleton is a central conduit for these forces. Zyxin and other LIM (Lin-11, Isl-1, Mec-3) domain proteins accumulate on strained F-actin through their LIM containing region (LCR), a process we term strain-selective binding. In cells, zyxin's LCR, which contains three tandem LIM domains, is necessary and sufficient for its force-dependent actin cytoskeleton localization, but the minimal unit required for strain-selective binding, in both the number of zyxin molecules and the number of LIM domains, is unknown. Here, we use an in vitro contractile network assay and multi-color TIRF microscopy to quantify the binding of full-length zyxin and LCR truncations to actomyosin bundles. Myosin II-dependent perturbations of F-actin bundles increase zyxin binding ~14-fold. Zyxin binds as clusters of 2-3 molecules with a median lifetime of ~80 s, independent of the number of molecules per cluster. The zyxin binding density is also independent of the number of actin filaments in a bundle. Each of zyxin's three LIM domains alone is sufficient for myosin-dependent binding, with binding density and lifetime similar to those of full length zyxin. The LIM domains differ in their tendency to cluster, which structure predictions suggest arises from hydrophobic LIM-LIM interfaces. Together, these data show that a single LIM domain is the minimal unit required for strain-selective binding of zyxin to F-actin. Zyxin binds rare, myosin-generated sites on actin filaments as clusters of 2-3 molecules, but neither clustering nor tandem LIM domains are required for this binding in vitro.

biophysics↗

Clustering conformations from molecular dynamics simulations by gradient flow segmentation of projected free energy landscapes

Molecular dynamics (MD) simulations sample a continuum of conformations whose populations can reveal distinct protein structural states and substates as well as different ligand binding states. Determining which conformations belong to each state remains a central challenge in interpreting protein structural dynamics and ligand binding. Building on our framework for constructing free-energy surfaces from principal component projections of MD trajectories, we treat these surfaces as images. We adapt a method originally developed to segment cell nuclei in three-dimensional microscopy images and use the basin structure of the free-energy surface as the basis for clustering. We follow gradient flow toward free-energy minima and group neighboring regions by comparing the free energy along their shared boundaries with that within each region. When both regions contain resolved minima, we also test how far apart their cores are relative to their internal variation. Mapping the regions associated with the minima back to the original trajectories recovers the conformational ensemble in each state or substate and allows us to select a representative conformation directly from the simulation. We applied the method to conformational changes in the SARS-CoV-2 spike protein and the leucine transporter LeuT, and to the binding conformations sampled by a small-molecule drug interacting with the PP2A scaffold PR65. By combining statistical mechanics with computer vision, this approach provides a structural interpretation of how sampled conformational populations are organized

biophysics↗