bioRxiv · 10.1101/2021.07.26.453893
Determination of best Raman spectroscopy spatial offsets for transcutaneous bone quality assessments in human hands
Abstract
Spatially offset Raman spectroscopy (SORS) is able to detect bone signal transcutaneously and could assist in predicting bone fracture risk. Criteria for optimal source-detector offsets for transcutaneous human measurements, however, are not well-established. Although larger offsets yield a higher percentage of bone signal, the absolute amount of bone signal decreases. Spectral unmixing into bone, adipose, and non-adipose components was employed to quantify changes in bone signal to noise ratio across a range of offsets, and optimal offsets for phalanx and metacarpal measurements were determined. The bone signal to noise ratio was maximized at offsets ranging from 4-6 mm.
Source connections
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Chen, K., Massie, C., Awad, H. A., Berger, A. J.. 2021-07-27. Determination of best Raman spectroscopy spatial offsets for transcutaneous bone quality assessments in human hands. https://doi.org/10.1101/2021.07.26.453893
Cite the original work for its findings. Save a collection to share your selection of sources.