Powering electronic implants by high frequency volume conduction: in human validation

Abstract

Wireless power transfer (WPT) is frequently used as an alternative to batteries to accomplish miniaturization in electronic medical implants (eMIs). However, established WPT methods require bulky parts within the implant or cumbersome external systems, hindering minimally invasive deployments and the development of networks of eMIs. As an alternative, we propose a WPT approach based on volume conduction of high frequency (HF) current bursts. These currents are applied through external textile electrodes and are collected by the eMIs through two electrodes at their opposite ends. This approach avoids bulky components to obtain power, making it possible to develop implants with a flexible threadlike conformation. In here we study in humans if HF (6.78 MHz) current bursts complying with safety standards and applied through two textile electrodes strapped around a limb can provide substantial powers from pairs of implanted electrodes. Time averaged electric powers obtained from needle electrodes (diameter = 0.4 mm, length = 3 mm, separation = 30 mm) inserted into arms and lower legs of five healthy participants were 5.9 {+/-} 0.7 mW and 2.4 {+/-} 0.3 mW respectively. We also report a procedure to characterize the coupling between the external system and the implants using two-port impedance models generated from medical images. The results demonstrate for the first time in humans that innocuous and imperceptible HF current bursts that flow through the tissues by volume conduction can be used to wirelessly power threadlike eMIs, overcoming the limitations of existing WPT methods in terms of minimal invasiveness and usability.

Source connections

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Minguillon, J., Tudela-Pi, M., Becerra-Fajardo, L., Perera, E., del-Ama, A. J., Gil-Agudo, A., Megia-Garcia, A., Garcia-Moreno, A., Ivorra, A.. 2021-03-16. Powering electronic implants by high frequency volume conduction: in human validation. https://doi.org/10.1101/2021.03.15.435404

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related preprints

Comparative study of chlorophyll measurement in Physcomitrium patens moss using a conventional microscope adapted for combined 2D+1D imaging and spectral analysis

Imaging spectroscopy often requires expensive and complex equipment. Here we show a simple procedure for attaching a standard miniature fiber spectrometer to a conventional microscope, allowing easy integration of 2D imaging with 1D high-resolution spectral measurements. This combination provides much of the benefit of a full imaging spectrometer without the large equipment investment, and we provide instructions for modifying microscopes to this setup and the present measurements of living cells that demonstrate their performance. Using this setup, we compare the quantitative measurement of chlorophyll concentration in Physcomitrium patens moss using color imaging and spectral sampling.

bioengineering

De novo designed single-domain antibodies protect against lethal cobra venom neurotoxicity in vivo

Generative protein design can now rapidly produce de novo binders with high affinity and functional activity against a wide range of targets, including lethal snake venom toxins. However, so far most reported successes rely on new-to-nature scaffolds with limited therapeutic precedent. Single-domain antibodies (VHHs) offer a clinically validated alternative scaffold that can bind and neutralize long-chain -neurotoxins, which are some of the most lethal components in snake venoms. Here we compare three recently established de novo design models with VHH-design capabilities (Germinal, RFantibody, and BoltzGen) for their ability to generate VHHs against the neurotoxin -cobratoxin from the monocled cobra (Naja kaouthia). Using standardized model inputs and evaluation criteria based on AlphaFold3 interface confidence (ipTM) and RMSD self-consistency, we find that Germinal was the only method to generate designs passing stringent in silico criteria for experimental testing. We therefore performed a larger Germinal design campaign employing three different VHH frameworks and experimentally validated 46 designs in vitro. Of these, 42 expressed as soluble proteins and we identified four binding hits derived from two of the three tested frameworks. Of the four binders, two lead candidates were further characterized and demonstrated high affinity (KDs of 4.1 nM and 10.8 nM), monomeric behavior and low polyreactivity, indicating favorable biophysical and developability properties, as well as functional toxin neutralization in vitro. To assess their therapeutic potential we investigated their ability to protect against -cobratoxin toxicity in vivo. Both candidates fully protected mice after -cobratoxin challenge, with 100% survival compared to a lethal control. One candidate also retained notable neutralization capacity against whole venom of Naja kaouthia with a survival of 56%, while the other protected 22% when tested in a rescue setting. Together, we demonstrate that de novo VHH design can generate high affinity single-domain antibodies with in vivo protection against lethal cobra venom neurotoxicity, and provide practical insights into method- and framework-dependent performance.

bioengineering

Simple Feedback for Complex Movement: Capturing Whole-Limb Reorganization during Single-IMU Gait Retraining

Clinical gait retraining typically relies on multi-sensor arrays and high-dimensional feedback displays, imposing setup and interpretation burdens that limit routine clinical deployment. We developed a single-IMU visual biofeedback system that delivers real-time feedback of Lower Limb Trajectory Error (LLTE), a composite kinematic error metric integrating knee position and shank angle across the stance phase. Twenty able-bodied adults walked on a treadmill under two visual biofeedback targets (flexed-knee, extended-knee) while receiving either corrected (n=10) or uncorrected (n=8) feedback, where the correction accounted for limb orientation at initial contact. LLTE and stance-phase knee kinematics adapted consistently under the flexed-knee target for both feedback groups, with feedback formulation moderating the temporal trajectory of change. Adaptation toward the extended-knee target was limited, likely because participants were already operating near terminal knee extension and because the scalar error metric provided limited directional information for correction. Ankle range of motion (ROM) changed significantly across the stance phase under both target conditions, while hip ROM did not. Multiscale multivariate sample entropy (MSMVSE) increased monotonically with time scale across all conditions, with no statistically distinguishable difference between corrected and uncorrected feedback. These results suggest that single-IMU LLTE biofeedback can modify gait mechanics and that adaptation was expressed across multiple lower-limb segments rather than through changes at a single joint.

bioengineering