Synthetic Photoplethysmography (PPG) of the radial artery through parallelized Monte Carlo and its correlation to Body Mass Index (BMI)
Abstract
Cardiovascular disease is one of the leading causes of death in the United States and obesity significantly increases the risk of cardiovascular disease. The measurement of blood pressure (BP) is critical in monitoring and managing cardiovascular disease hence new wearable devices are being developed to make the BP metric mode accessible to physicians and patients. Several wearables utilize photoplethysmography from the wrist vasculature to derive BP assessment although many of these devices are still at the experimental stage. With the ultimate goal of supporting instrument development, we have developed a model the photoplethysmographic waveform derived from the radial artery at the volar surface of the wrist. To do so we have utilized the relation between vessel biomechanics through Finite Element Method and Monte Carlo light transport model. The model shows similar features to that seen in PPG waveform captured using an off the shelf device. We observe the influence of body mass index (BMI) on the PPG signal. A degradation the PPG signal of up to 40% in AC to DC signal ratio was thus observed.
Source connections
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Boonya-Ananta, T., Rodriguez, A. J., Ajmal, A., Hansen, A. K., Hutcheson, J. D., Ramella-Roman, J. C.. 2020-07-24. Synthetic Photoplethysmography (PPG) of the radial artery through parallelized Monte Carlo and its correlation to Body Mass Index (BMI). https://doi.org/10.1101/2020.07.22.214015
Cite the original work for its findings. Save a collection to share your selection of sources.