bioRxiv · 10.1101/2020.04.21.052530
One-pot Detection of COVID-19 with Real-time Reverse-transcription Loop-mediated Isothermal Amplification (RT-LAMP) Assay and Visual RT-LAMP Assay
Abstract
BackgroundRapid and reliable diagnostic assays were critical for prevention and control of the coronavirus pneumonia caused by COVID-19. ObjectiveThis study was to establish one-pot real-time reverse-transcription loop-mediated isothermal amplification (RT-LAMP) assay and one-pot visual RT-LAMP assay for the detection of COVID-19. MethodsSix specific LAMP primers targeting the N gene of COVID-19 were designed, the RT-LAMP reaction system was optimized with plasmid pUC57 containing N gene sequence, the detection limit was determined with a serial dilution of the plasmid pUC57 containing N gene sequence, and the one-pot real-time RT-LAMP assay and one-pot visual RT-LAMP assay for the detection of COVID-19 were established. ResultsOur results showed that the one-pot RT-LAMP assays can detect COVID-19 with a limit of [≥] 6 copies per l-1 of pUC57 containing N gene sequence. ConclusionThis study provides rapid, reliable and sensitive tools for facilitating preliminary and cost-effective prevention and control of COVID-19.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Wang, D.. 2020-04-22. One-pot Detection of COVID-19 with Real-time Reverse-transcription Loop-mediated Isothermal Amplification (RT-LAMP) Assay and Visual RT-LAMP Assay. https://doi.org/10.1101/2020.04.21.052530
Cite the original work for its findings. Save a collection to share your selection of sources.