Programming Continuous, High Framerate, and Long-Duration Data Acquisitions on the Verasonics Ultrasound System

The ultrasound community has seen a rise in popularity of highly demanding imaging acquisition sequences, such as those for super-resolution ultrasound localization microscopy and functional ultrasound imaging, which push the boundary of ultrasound system hardware limits. However, subtle errors in sequence programming can lead to corrupted data, inaccurate timing, and slow data throughput, which impact the performance of these modalities. In this article, we detail the fundamentals of the hardware and software sequencers for the Verasonics Vantage/NXT research ultrasound scanner, highlighting some common pitfalls in sequence design and ways to overcome them. We include example imaging sequences for different data acquisition goals, and detail recommended ways to test these sequences to ensure that they are meeting the expected performance.